• Home
  • Biopharmaceutical Research Services
  • Multi-Omics Services
  • Support
  • /assets/images/icon/icon-email-2.png

    Email:

    info@MtoZ-Biolabs.com

    Why Is There Only a Solvent Peak but No Analyte Peak in GC-MS Standard Run

      In the GC-MS analysis of a standard sample, the observation of a solvent peak without the presence of an analyte peak may result from several potential factors, as outlined below:

       

      1. Sample Preparation Issue

      Verify that the sample preparation procedure has been correctly performed. Operational errors during the preparation of the standard may occur, for example, the analyte may not have been added to the solvent or was added in an insufficient amount.

       

      2. Syringe Issue

      Ensure that the syringe is intact, free of blockages, and uncontaminated. A faulty or contaminated syringe may hinder the complete transfer of the sample into the chromatographic column.

       

      3. Injection Volume Issue

      Confirm that the injection volume is appropriate. An excessively small injection volume may prevent the analyte peak from being detected.

       

      4. Column Issue

      Check whether the chromatographic column is properly installed, with tightly connected inlet and outlet ends. Contamination or internal damage to the column may also interfere with the detection of the analyte peak.

       

      5. Chromatographic Condition Issue

      Review the GC settings, such as the temperature program and carrier gas flow rate. Suboptimal chromatographic conditions may lead to poor separation or failure to detect the analyte peak.

       

      6. MS Parameter Issue

      Verify the mass spectrometry settings, including the electron ionization (EI) voltage, ion source temperature, and detector gain. Improper parameter settings may either inhibit ionization or prevent the successful detection of the analyte ions.

       

      7. Selected Ion Monitoring (SIM) Mode Issue

      If the SIM mode is employed, ensure that the correct mass-to-charge (m/z) values have been selected. An incorrect choice of m/z values may result in the failure to detect the analyte peak.

       

      8. Detection limit issue

      Assess whether the analyte concentration falls below the instrument’s detection limit. If the analyte concentration is too low, the peak may not be observable.

       

      To resolve this issue, systematically investigate and rule out the potential causes listed above, making appropriate optimizations based on the specific analytical context. If the analyte peak remains undetectable, it may be necessary to seek expert consultation or consider alternative analytical techniques.

       

      MtoZ Biolabs, an integrated chromatography and mass spectrometry (MS) services provider. 

      Related Services

      Metabolomics Analysis Service

    Submit Inquiry
    Name *
    Email Address *
    Phone Number
    Inquiry Project
    Project Description *

     

    How to order?


    /assets/images/icon/icon-message.png

    Submit Inquiry

    /assets/images/icon/icon-return.png